User's Guide ExStik® EC500 pH/Conductivity / TDS / Salinity / Temperature Meter EC500

Pagina 2 - Getting Started

EC500 V2.1 7/13/05 10 Operational Matrix Function / Resulting Action Power Status Mode Setting Required Key Press Sequence On/Off On or Off Any

Pagina 3 - Specifications

EC500 V2.1 7/13/05 11 Maintenance Battery Replacement 1. Twist off the battery compartment cap 2. Holding the battery housing in place with a fin

Pagina 4 - Meter Description

EC500 V2.1 7/13/05 12 Cleaning Recommendations When cleaning the probe, take care not to scratch or damage the sensing surface or the platinized ele

Pagina 5 - Measurement:

EC500 V2.1 7/13/05 13  Support line (781) 890-7440 Technical support: Extension 200; E-mail: [email protected] Repair & Returns: Extension 210

Pagina 6 - Storing Readings

EC500 V2.1 7/13/05 2 Introduction Congratulations on your purchase of the ExStik® EC500 pH/Conductivity/Total Dissolved Solids (TDS) / Salinity mete

Pagina 7 - Recalling Stored Readings

EC500 V2.1 7/13/05 3 Specifications Display 2000 count LCD with Bargraph pH Range 0.00 to 14.00 pH Accuracy ±0.01 pH typical pH ATC Range 0°C to

Pagina 8 - RENEW Display

EC500 V2.1 7/13/05 4 EC5007RENEW6pHMeter Description Front Panel Description 1. Battery compartment cover 2. LCD Display 3. MODE/HOLD button 4.

Pagina 9 - Calibration - Conductivity

EC500 V2.1 7/13/05 5 8888ConductivityTDSSalinityPHLMHpH(ppm)(mg/l)mgLMeasurement Procedure Sample Preparation: 1. For Conductivity, TDS or Salinity

Pagina 10 - EC500 V2.1 7/13/05

EC500 V2.1 7/13/05 6 Changing Measurement Function The meter can be set to measure pH, Conductivity, TDS or Salinity. To change the mode: 1. Press

Pagina 11 - Maintenance

EC500 V2.1 7/13/05 7 Recalling Stored Readings 1. Press the CAL/RECALL button and then press the MODE/HOLD button. A location number (1 through 25)

Pagina 12 - Cleaning Recommendations

EC500 V2.1 7/13/05 8 Calibration - pH (1, 2, or 3 points) 1. Place the electrode into a buffer solution (4, 7, or 10). Press and hold the CAL/RECA

Pagina 13 - Troubleshooting

EC500 V2.1 7/13/05 9 Calibration - Conductivity Meter accuracy verification should be performed on a periodic basis. Once per month is the recommend