Multi 3430 Operation
41
ba75865e11 02/2011
Note
You can prematurely terminate the Stability control function manually
with <MENU/ENTER> at any time. If the Stability control function is pre-
maturely terminated, the current measurement data are output to the
interface without the AutoRead info.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
3 Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back
to the previous indication.
Measured
parameter
Time interval Stability in the time interval
pH value 15 seconds
Δ pH: Better than 0.01
Temperature 15 seconds
Δ T (° C): Better than 0.02